Note: Only use the PSU supplied with the device for safety purposes. Withings warranty does not cover defects due to the use of spare parts not provided by Withings in the original packaging.
ペースメーカーを装着している場合、Sleep Analyzerを使用するべきですか?
はい。ペースメーカーやその他の内蔵医療機器を装着している場合も、Sleepをご使用いただけます。
CPAPマシンを持っている場合、Sleep Analyzerを使用すべきですか?
We are currently investigating the use of the Withings Sleep Analyzer with different CPAP devices. Sleep has not been yet tested for use during CPAP treatment. A scientific protocol is currently ongoing in CPAP users to compare AHI (Apnea-Hypopnea Index) given by CPAP devices and the WSA (Wake-Sleep Algorithm).
妊娠中にSleep Analyzerを使用するべきですか?
はい。妊娠中の女性がSleep Analyzerを使用することに関連する既知のリスクはございません。
子どもの睡眠を監視するためにSleep Analyzerを使用するべきですか?
Sleep Analyzerは大人(18歳以上)でテストされています。残念ながら、お子様の睡眠の正確さを保証するものではございません。
Should I use Sleep Analyzer if the device becomes warm?
Yes, it is normal for the Sleep Analyzer to become warm, particularly at night. This is in part the result of the device being in between the mattress and bed frame, making it difficult for heat to dissipate. Keep in mind, however, that the Sleep Analyzer was designed to operate under these conditions.
Are there any contraindications to the use of the Sleep Apnea Detector of Sleep Analyzer?
Sleep Apnea Detector is not indicated for use on patients with the following conditions:
- Arrhythmia, as the device also analyzes changes in the heart frequency to detect apnea and hypopnea events. Since arrhythmias are often characterized by irregular heartbeats, these conditions may mislead the device.
- Cardiac insufficiency (heart failure), as this condition may be associated with irregular heart rhythm and breathing which may mislead the device and has not been tested.
-
Chronic obstructive pulmonary disease, such as nocturnal asthma, has a distinctive breathing pattern with prolonged expirations, which the device may confuse with apnea or hypopnea breathing patterns, incorrectly increasing the AHI.
Chronic hypoventilation observed in chronic obstructive pulmonary disease or neuromuscular diseases (Charcot’s disease) has a different breathing pattern than apneas, with prolonged shallow amplitude of breathing.
Other conditions may also negatively affect the performance of the device. These include disorders such as periodic limb movements in sleep and restless legs syndrome, which could create interfering signals, or non-respiratory sleep disorders that result in fragmented sleep.